Accurate DNA, RNA & protein quantification in less than 3 seconds

The Colibri+ takes microvolume quantification of DNA, RNA and protein samples to a whole new level. The system has been designed to help you accelerate your research combining ultra-fast measurement in less than 3 seconds with intuitive 7-inch touchscreen stand-alone operation.


Broad detection range
    - 2 – 20.000 ng/µL for dsDNA.
    - 0.06 – 600 mg/mL for BSA.

Reliable results
    - CV of absorbance  < 1%.

Fast measurement
    - Read your samples in less than 3 seconds.

Easy pipetting of the sample
    - The Sample Guide Light enables easy sample application and helps you to detect even the smallest air bubbles. As a result, the quality of the measurement is significantly improved.

Accurate and reproducible measurements
    - Hydrophobic coating of sample window: The hydrophobic coating of the sample window supports the formation of a measuring column from the sample liquid after closing the detector arm.
    - Shock-absorbing detector arm: The shock-absorbing detector arm reduces shocks during its closure. This supports accurate measurement of samples even as little as 1 µl with a CV of absorbance < 1%.

Preprogrammed applications
    - A wide variety of preprogrammed protocols simplifies operation of the system.
    - Easily create your own protocols.

Standalone operation
    - The intuitive 7‘’ colour touchscreen operation enables you to analyze your samples with the touch of a button.
    - 32 GB of onboard memory.
    - Create more than 500 user folders, which can be password protected.
    - Saves precious lab space in combination with its small footprint of just 206 x 333 x 166 mm (W x D x H).

Specification:
    -Minimum Sample Size    1 µL
    -Detection Range dsDNA    2 - 20,000 ng/µL
    -Detection Range BSA    0.06 - 600 mg/mL
    -Path Length    0.5 and 0.05 mm
    -Measurement Time    < 3 sec
    -Display    7" glove compatible colour LCD touchscreen

    - Light Source    Pulsed Xenon flash lamp
    - Detector    2,048 CMOS
    - Wavelength Range    190 - 1,000 nm
    - Wavelength Accuracy    1.3 nm
    - Absorbance Accuracy    3.0% (at 0.75 A at 300 nm)
    - Absorbance Range (10 mm equivalent)    0.04 - 400 A
    - Spectral Resolution    1.5 nm (FWHM at Hg 253.7 nm)
    - Dimensions (W x D x H)  206 x 333 x 166 mm
    - Weight   3.5 kg (7.8 lb)

Are you performing real-time PCR, sequencing / NGS, or cloning experiments? Then the Colibri+ spectrophotometer will help you advance your research by supporting a broad range of applications:

    - Determination of nucleic acid concentration and purity of nucleic acid samples without dilution
    - Purified protein analysis (A280)
    - Bradford, BCA and Lowry assays for protein quantification
    - Cell density OD600 measurements
    - General UV-VIS spectrophotometry